Skip to main content

Table 1 Activity concentrations and SBRs for all phantom acquisitions as determined with the dose calibrator

From: Image quality assessment along the one metre axial field-of-view of the total-body Biograph Vision Quadra PET/CT system for 18F-FDG

SBR (calculated)

Axial position (mm)

Sphere activity concentration (kBq/ml)

Background activity concentration (kBq/ml)

Fraction of rebinned acquisition time used for reconstruction

8:1 (8.0:1)

0

26.02

3.26

0.77

250

23.66

2.97

0.84

450

21.66

2.72

0.92

505

19.95

2.50

1.00

4:1 (3.9:1)

0

12.99

3.30

0.76

250

11.81

3.00

0.84

450

10.88

2.76

0.91

505

9.90

2.51

1.00

2:1 (2.0:1)

0

6.48

3.29

0.78

250

6.01

3.05

0.84

450

5.47

2.77

0.92

505

5.04

2.56

1.00