Skip to main content

Table 2 Errors and uncertainties in quantification using the proposed method for four lesions in the anthropomorphic phantom

From: Quantification and dosimetry of small volumes including associated uncertainty estimation

Lesion location

Error (MBq)

Error (%)

Uncertainty (%)

A

0.34

− 8.9

2.9

B

0.23

− 5.1

1.2

C

0.41

− 11.6

9.6

D

0.04

− 0.8

5.0