Table 3 Root mean square (rms) error in units of ml/min/1.73m2 between single-sample GFR and reference GFR measurements

From: Tailoring the sampling time of single-sample GFR measurement according to expected renal function: a multisite audit

Reference GFR mL/min/1.73m2

Number of patient studies

Sampling time (h)

2

3

4

6

30–40

2

9.62

8.14

7.02

7.18

40–50

3

4.27

3.62

5.53

9.49

50–60

15

8.67

9.84

9.88

11.89

60–70

37

8.45

9.49

10.66

13.51

70–80

49

5.81

6.49

7.65

11.91

80–90

61

7.00

7.71

8.86

13.41

90–100

67

5.17

5.78

7.11

11.58

100–110

69

5.07

5.85

7.39

11.82

110–120

56

5.61

6.42

6.79

8.60

120+

59

7.33

6.86

7.43

8.99