Table 4 Means and standard deviations of image quality parameters determined for NEMA PET body phantom measurements

From: Comparison of image quality and spatial resolution between 18F, 68Ga, and 64Cu phantom measurements using a digital Biograph Vision PET/CT

 

Contrast: ≈ 4:1: normal weight setup

Contrast: ≈ 4:1: obese setup

Contrast: ≈ 8:1: normal weight setup

Contrast: ≈ 8:1: obese setup

RCmean

18F-FDG

1.06 ± 0.01

1.08 ± 0.02

1.00 ± 0.04

1.02 ± 0.05

68Ga-HCl

0.97 ± 0.09

1.03 ± 0.08

0.95 ± 0.06

0.96 ± 0.05

64Cu-HCl

1.05 ± 0.03

1.07 ± 0.02

1.10 ± 0.03

1.12 ± 0.03

RCmax

18F-FDG

1.38 ± 0.05

1.45 ± 0.04

1.33 ± 0.09

1.37 ± 0.11

68Ga-HCl

1.28 ± 0.13

1.41 ± 0.15

1.23 ± 0.04

1.28 ± 0.07

64Cu-HCl

1.38 ± 0.06

1.43 ± 0.05

1.42 ± 0.09

1.51 ± 0.06

RCpeak

18F-FDG

1.12 ± 0.23

1.10 ± 0.25

1.00 ± 0.33

1.03 ± 0.34

68Ga-HCl

1.01 ± 0.13

1.08 ± 0.16

1.00 ± 0.16

0.99 ± 0.21

64Cu-HCl

1.10 ± 0.22

1.10 ± 0.27

1.09 ± 0.33

1.13 ± 0.34

CRC (%)

18F-FDG

100.99 ± 1.72

101.57 ± 2.99

96.46 ± 4.07

97.86 ± 5.24

68Ga-HCl

81.99 ± 10.88

76.48 ± 8.32

80.98 ± 5.75

91.20 ± 5.06

64Cu-HCl

98.97 ± 3.67

98.79 ± 2.18

103.90 ± 3.32

106.09 ± 3.09

CNR

18F-FDG

34.26 ± 0.58

24.60 ± 0.72

86.55 ± 3.65

64.19 ± 3.44

68Ga-HCl

27.83 ± 3.69

19.36 ± 2.11

70.48 ± 5.00

61.10 ± 3.39

64Cu-HCl

46.86 ± 1.74

33.24 ± 0.73

76.40 ± 2.44

57.87 ± 1.69

CoVBG (%)

18F-FDG

8.96 ± 0.33

12.55 ± 0.15

9.72 ± 0.21

13.29 ± 0.77

68Ga-HCl

9.60 ± 0.40

12.87 ± 0.41

9.67 ± 0.43

12.56 ± 0.81

64Cu-HCl

9.30 ± 0.64

13.09 ± 0.63

9.68 ± 0.69

13.04 ± 0.77

Lung count error [%]

18F-FDG

3.82

5.42

4.19

5.22

68Ga-HCl

7.85

13.48

8.23

5.90

64Cu-HCl

3.72

5.11

3.76

4.95

  1. Means and standard deviations of the mean, maximum and peak recovery coefficients (RCmean, RCmax, and RCpeak, respectively), percent contrast between each sphere and the background (contrast recovery coefficient (CRC)), contrast–noise ratio (CNR), percent background variability (CoVBG) and the relative count error in the lung insert (lung count error). Except for the relative count error in the lung insert, each parameter is presented as the mean ± standard deviation of the six spheres for PET/CT measurements of the NEMA PET body phantom filled with 18F-FDG, 68Ga-HCl, or 64Cu-HCl with sphere-to-background activity concentration ratios of approximately 4:1 or 8:1 and with (obese setup) or without (normal weight setup) the simulation of an obese patient by applying cooling packs around the phantom