Fig. 7 | EJNMMI Physics

Fig. 7

From: Monte Carlo modelling of a compact CZT-based gamma camera with application to 177Lu imaging

Fig. 7

Simulated and measured system sensitivities as function of the source-collimator distance for \(^{177}{\mathrm {Lu}}\) for the LEHR and MEGP collimators and energy windows positioned over 55 keV, 113 keV and 208 keV. Simulated data are connected with straight lines, for measured data a function is fitted. Simulated sensitivities are separated into sub-components based on the emission energy and events in the photon transport, including collimated primary (geom.  primary), phantom scatter (phan.  sca.), collimator penetration (pen.), collimator scatter (col.  sca.) and collimator x-rays (x-rays). Components with a low contribution are grouped together (other)