Table 1 EARL1 and EARL2 RC limits. *At the time of this work, RCmax for total-body scanners was under investigation. ** RCpeak limits were under revision. Table

From: EARL compliance measurements on the biograph vision Quadra PET/CT system with a long axial field of view

NEMA IEC phantom spheres

18F standard 1 RC limits

18F standard 2 RC limits

Sphere i

Diameter (mm)

Volume (mL)

RCmax

RCmean

RCmax*

RCmean

RCpeak**

1

37

26.52

0.95–1.16

0.76–0.89

1.05–1.29

0.85–1.00

0.90–1.10

2

28

11.49

0.91–1.13

0.72–0.85

1.01–1.26

0.82–0.97

0.90–1.10

3

22

5.57

0.83–1.09

0.63–0.78

1.01–1.32

0.80–0.99

0.90–1.10

4

17

2.57

0.73–1.01

0.57–0.73

1.00–1.38

0.76–0.97

0.75–0.99

5

13

1.15

0.59–0.85

0.44–0.60

0.85–1.22

0.63–0.86

0.45–0.70

6

10

0.52

0.34–0.57

0.27–0.43

0.52–0.88

0.39–0.61

0.27–0.41