Table 9 Comparison of the recovery coeficient (RC) for the 3mm diameter rod, uniformity and Spill-Over Ratio (SOR) of different preclinical PET scanners. Data are taken from  [5,6,7,8,9,10,11,12,13,14,15, 38]

From: NEMA characterization of the SAFIR prototype PET insert

System

Recon. method

Corrected for AC/SCb

RC

STD (%)

SOR (%)

Uniform region

Water chamber

Air chamber

SAFIR

3D MLEM

yes/yes

0.49

3.0

17.3

18.5

Hyperion IID

3D MLEM

yes/yes

0.91

3.7

5.4

6.3

NanoScan

penalized MLEM

yes/yes

0.9

3.5

6.2

5.8

MuPET

FBP3DRP

yes/no

0.75

6.5

9

5

Inveon

FORE + 2D FBP

yes/yes

0.72

5.3

1.7

−0.6

IRIS

OSEM

yes/no

0.73

7

11

11

ClearPET

3D OSEMa

no/no

0.42

10.9

36.9

26.7

Mosaic HP

3D RAMLAa

yes/yes

0.56

5.1

6.3

2.7

LabPET 8TM

2D MLEM

no/no

0.58

7.0

20

11

microPET R4

FOREa + 2D FBP

yes/no

0.60

4.5

6.2

4.6

Xtrim-PET

FORE + 2D OSEM

no/no

0.68

3.8

25

35

Bruker

3D MLEM

yes/no

0.91

4.5

6.2

4.6

  1. aFORE = Fourier Rebinning; RAMLA = Row-Action Maximum-Likelihood Algorithm; OSEM = Ordered Subset Expectation Maximization
  2. b AC = Attenuation Correction; SC = Scatter Correction