Skip to main content

Table 3 Comparison of the peak sensitivity of different preclinical PET scanners. Data are taken from [5,6,7,8,9,10,11,12,13,14,15]

From: NEMA characterization of the SAFIR prototype PET insert

System

Energy win. (keV)

Time win. (ns)

Axial FOV (mm)

Sensitivity (%)

SAFIR

391–601

0.5

36

1.06

SAFIR

250–650

0.5

36

2.57

Hyperion IID

250–625

200

96.7

4.0

NanoScan

250–750

5

94

8.4

MuPET

350–650

3.4

116

6.35

Inveon

350–625

3.4

127

6.72

IRIS

250–750

5.2

95

8

ClearPET

250–650

12

110

1.87

Mosaic HP

385–665

7

119

1.77

LabPET 8TM

250–650

20

75

1.33

microPET R4

350–650

6

78

2.4

Xtrim-PET

250–650

10

50.3

2.99

Bruker

150

11.0