Table 1 FWHM and FWTM at axial center resolutions at different radial positions

From: Performance evaluation of the IRIS XL-220 PET/CT system, a new camera dedicated to non-human primates

Radial distance (mm) Tangential (mm)
Rodent/primate
Radial (mm)
Rodent/primate
Axial (mm)
Rodent/primate
FWHM FWTM FWHM FWTM FWHM FWTM
0 0.96/1.22 2.26/2.22 1.08/1.07 2.34/1.96 1.52/1.1 2.44/2.00
5 0.93/1.19 2.21/2.16 1.00/1.01 2.27/1.84 1.58/1.09 2.45/1.99
10 1.03/1.09 2.29/1.98 1.09/1.08 2.38/1.96 1.67/1.12 2.46/2.04
15 1.00/1.06 2.34/1.93 1.10/1.29 2.61/2.36 1.74/1.09 2.46/1.99
25 1.14/1.39 2.68/2.54 1.36/1.52 3.11/2.77 1.47/1.05 2.42/1.92
50 1.40 2.55 1.90 3.47 0.92 1.67
75 1.51 2.75 2.11 3.85 0.91 1.66
  1. Slice thickness (mm): 0.84