Skip to main content

Table 3 SUVmax and SUVmean for ϕ 22-mm sphere obtained using two gamma cameras: Discovery NM/CT 670 and NM/CT 850, in the two medical departments. The radioactive decay corrections were included in the calculations

From: What validation tests can be done by the clinical medical physicist while waiting for the standardization of quantitative SPECT/CT imaging?

Gamma camera

Sphere volume Q.Metrix

[ml]

Sphere activity Q.Metrix

[MBq]

SUVmean

[g/ml]

SUVmax

[g/ml]

SUVmin

[g/ml]

Discovery NM/CT 670

Department 1

5.8

8.17

52.40

113.30

7.18

NM/CT 850

Department 2

5.8

6.47

59.05

112.62

6.56