Table 2 Parameters of image quality obtained for the 22-mm-diameter sphere and the background of the NEMA/IEC NU2 Body Phantom with two gamma cameras, Discovery NM/CT 670 and NM/CT 850

From: What validation tests can be done by the clinical medical physicist while waiting for the standardization of quantitative SPECT/CT imaging?

 

Parameter

Activity concentration in the hot sphere

Ac,sphere (22 mm) [MBq/ml] (measured with a meter)

Activity concentration in the hot sphere measured with Q.Metrix

ac,sphere (22 mm) [MBq/ml]

(measured with Q.Metrix)

Hot contrast QH (%)

22-mm sphere

(measured in the image)

Activity concentration in the background

Ac,bg [MBq/ml]

(measured with a meter)

Activity concentration in the background

ac,bg [MBq/ml]

(measured with Q.Metrix, mean for 5 ROIs)

Discovery NM/CT 670

Department 1

0.32

0.42

71

0.04

0.07

NM/CT 850

Department 2

0.32

0.38

76

0.04

0.06