Table 1 Background calibration factor (Bg. cal), coefficient of variation (COV), total activity deviation (ΔAtot), background activity concentration (Ac,bg) measured with the activity meter and background activity concentration (ac,bg)—average of the 5 ROIs measured in the image with Q.Metrix, for two SPECT/CT acquisitions of the cylindrical uniform phantom in two nuclear medicine departments

From: What validation tests can be done by the clinical medical physicist while waiting for the standardization of quantitative SPECT/CT imaging?

Gamma camera, medical center

Coefficient

Background calibration factor

Bg. cal

Coefficient of variation COV (%)

Total activity deviation

ΔAtot

(%)

Activity concentration in the background

Ac,bg

(MBq/ml)

(measured with an activity meter)

Activity concentration in the background

ac,bg (MBq/ml)

(measured in the image with Q.Metrix, average of the 5 ROIs)

Discovery 670

Department 1

1.192

4

− 1.02

0.052

0.062

NM/CT 850

Department 2

1.135

10

− 0.45

0.052

0.059