From: Harmonization of PET image reconstruction parameters in simultaneous PET/MRI
Metric | Â | Optimized high CRC | Intermediate | Lowest RMSD |
---|---|---|---|---|
CRC-Mean | Siemens mMR | 4 it – 3 mm + PSF | 4 it – 3 mm | 4 it – 6 mm |
 | GE Signa | 2 it – 3 mm + IR | 2 it – 5 mm + IR | 2 it – 7 mm |
RMSD | Â | 2.79 | 2.07 | 1.82 |
CRC- Max | Siemens mMR | 4 it – 3 mm + PSF | 3 it – 4 mm + PSF | 4 it – 5 mm |
 | GE Signa | 2 it – 3 mm + IR | 2 it – 5 mm + IR | 4 it – 6 mm |
RMSD | Â | 5.83 | 4.89 | 3.57 |
CRC-Peak | Siemens mMR | 4 it – 3 mm + PSF | 3 it – 3 mm | 4 it – 7 mm + PSF |
 | GE Signa | 4 it – 4 mm + IR | 4 it – 6 mm + IR | 4 it – 7 mm |
RMSD | Â | 3.23 | 3.05 | 2.91 |