Skip to main content

Table 6 Uncertainty budget for renal cortex activity ratios (ASPECT: SPECT-based activity, ACalibrator: radionuclide calibrator-based activity, R(α) / R(β) / R(V): uncertainty components introduced by the recovery originating from α, β and V, respectively)

From: A multicentre and multi-national evaluation of the accuracy of quantitative Lu-177 SPECT/CT imaging performed within the MRTDosimetry project

Setup

ASPECT/ ACalibrator

Uncertainty components (%)

Combined standard uncertainty (%)

Time

Counts

R(α)

R(β)

R(V)

ACalibrator

ICF

S2a

0.88

0.03

0.03

1.99

4.47

4.08

2.83

3.13

7.64

S2b

0.75

0.03

0.05

2.11

4.77

2.23

2.83

3.13

7.07

S2c

0.77

0.03

0.05

0.79

2.15

1.25

2.83

3.13

4.96

S4

0.77

0.03

0.06

2.51

1.62

8.03

0.52

0.52

8.60

S7

0.78

0.03

0.03

1.15

3.10

3.20

4.24

4.24

7.56