Skip to main content

Table 5 Spatial resolution and sensitivity under different conditions

From: Design and system evaluation of a dual-panel portable PET (DP-PET)

PET performance on position (0, 0 ,0) mm MR idle MR pulsing
Spatial resolution on the x direction (mm) 2.29 ± 0.01 2.29 ± 0.03
Spatial resolution on the y direction (mm) 2.36 ± 0.01 2.36 ± 0.03
Spatial resolution on the z direction (mm) 7.30 ± 0.13 7.32 ± 0.12
Sensitivity 3.59% ± 0.03% 3.58% ± 0.02%