Skip to main content

Table 4 SNR and image uniformity under different conditions

From: Design and system evaluation of a dual-panel portable PET (DP-PET)

Conditions SNR Image uniformity
No DP-PET with Dixon in-phase 5.04 ± 0.00 83.74% ± 0.04%
DP-PET inserted with Dixon in-phase 4.85 ± 0.02 82.90% ± 0.05%
No DP-PET with Dixon out-phase 4.79 ± 0.01 83.12% ± 0.03%
DP-PET inserted with Dixon out-phase 4.70 ± 0.04 82.38% ± 0.13%