Table 3 Sensitivity at different positions

From: Design and system evaluation of a dual-panel portable PET (DP-PET)

Source position (x, y, z) mm Sensitivity
(0, 0, 0) 3.59% ± 0.03%
(0, 10, 0) 3.36% ± 0.02%
(0, 20, 0) 3.01% ± 0.01%
(0, 30, 0) 2.73% ± 0.01%
(0, 40, 0) 2.27% ± 0.01%
(0, 0, 10) 3.28% ± 0.02%
(0, 0, 20) 2.84% ± 0.02%
(0, 0, 30) 2.52% ± 0.03%
(0, 0, 40) 2.30% ± 0.01%