Table 1 Activities and activity concentrations in the cylinders used for calibration and verification

From: Tissue dose estimation after extravasation of 177Lu-DOTATATE

Acquisition Calibration-cylinder 1 Calibration-cylinder 2 Verification-cylinder
A (MBq) AC (MBq/mL) A (MBq) AC (MBq/mL) A (MBq) AC (MBq/mL)
1 2275 ± 59 7.51 ± 0.20 1136 ± 30 3.54 ± 0.09 471 ± 12 3.14 ± 0.08
2 1700 ± 44 5.61 ± 0.15 849 ± 22 2.65 ± 0.07 354 ± 9 2.36 ± 0.06
3 808 ± 21 2.67 ± 0.07 403 ± 10 1.26 ± 0.03 167 ± 4 1.12 ± 0.03
4 388 ± 10 1.28 ± 0.03 194 ± 5 0.60 ± 0.02 80 ± 2 0.54 ± 0.01
5 186 ± 5 0.61 ± 0.02 93 ± 2 0.29 ± 0.01 39 ± 1 0.26 ± 0.01
6 90 ± 2 0.30 ± 0.01 45 ± 1 0.140 ± 0.004 19 ± 0.5 0.12 ± 0.003