Skip to main content

Table 2 Sensitivity measured with different source geometries on Anger and CzT cameras with and without focus mode

From: Evaluation of a new multipurpose whole-body CzT-based camera: comparison with a dual-head Anger camera and first clinical images

Sensitivity in counts·s−1·MBq−1Anger cameraCzT camera without focusCzT camera with focus
Point source in air143.61236.681159.33
Point source in head phantom56.9773.42342.40
Point source in body phantom35.9439.05207.70
Head phantom72.69107.18na
Body phantom52.7565.79na