Skip to main content

Table 1 Percent FWHM measured on CzT and Anger cameras for 6 energy peaks. Results are expressed in mean (SD)

From: Evaluation of a new multipurpose whole-body CzT-based camera: comparison with a dual-head Anger camera and first clinical images

Radioisotope

201Tl

99mTc

123I

111In

Energy peak (keV)

70

167

140

159

171

245

% FWHM Anger camera

14.08 (0.19)

11.98 (1.66)

9.21 (0.07)

10.24 (0.79)

10.78 (0.03)

9.03 (0.35)

% FWHM CzT camera

5.77 (1.30)

4.67 (1.13)

5.46 (0.59)

5.33 (0.62)

6.27 (1.08)

4.5 (0.75)