Fig. 5 | EJNMMI Physics

Fig. 5

From: Simultaneous 166Ho/99mTc dual-isotope SPECT with Monte Carlo-based downscatter correction for automatic liver dosimetry in radioembolization

Fig. 5

Recovered 166Ho activity as a percentage of the known injected activity for 166Ho filled spheres in the NEMA image quality phantom for various background concentrations of 99mTc (columns). The top row shows the recovered 166Ho activity as a function of the number of UMCS-OSEM iterations (8 subsets per iteration) in VOIs matching actual sphere sizes (solid lines) and VOIs with diameters increased by 20 mm (dotted lines). The bottom row shows the recovered 166Ho activity after 20 iterations as a function of increased VOI diameter (i.e., in addition to the actual sphere size). For clarity, results of only 3 out of the 6 NEMA IQ spheres are shown