Skip to main content
Fig. 4 | EJNMMI Physics

Fig. 4

From: Explicit measurement of multi-tracer arterial input function for PET imaging using blood sampling spectroscopy

Fig. 4

a Calibration profiles obtained in the well counter for 300-μl pure 18F (red squares) and 68Ga (blue circles) samples with different activity values using the full energy window (W200–2000). Each dataset was fitted to a straight line with y-intercept forced to be 0 obtaining the calibration factors εF = 0.335 cps Bq−1 and εGa = 0.371 cps Bq−1 at this volume. b Variation of calibration factors with the sample volume for 18F (εF) and 68Ga (εGa)