Fig. 2 | EJNMMI Physics

Fig. 2

From: Cross-validation study between the HRRT and the PET component of the SIGNA PET/MRI system with focus on neuroimaging

Fig. 2

RCs and %BG variability as a function of sphere diameter for the contrast phantom (PET/MR: RCs (a), %BG variability (b); HRRT: RCs (c), %BG variability (d)). PET/MR data were reconstructed with TOF, TOF with filter, TOF with PSF, and TOF with PSF and filter. Two iterations and 28 subsets were used, respectively (GE recommendation for phantom data). HRRT data were reconstructed without any filters (native), with a 2-mm Gaussian filter (standard in-house reconstruction for human data) using 6 iterations and 16 subsets, respectively, with PSF, and with PSF correction and 2-mm filter using 10 iterations and 16 subsets, respectively. Note the gap between spheres marks cold vs. hot spheres