Skip to main content

Table 5 Bland Altman analysis, Pearson’s correlation, and Lin concordance coefficients, of mean absorbed doses delivered to both tumoral liver and individual tumors volumes among the studied dosimetry methods (PM, MTPM, LDM, and DPK)

From: Impact of the dosimetry approach on the resulting 90Y radioembolization planned absorbed doses based on 99mTc-MAA SPECT-CT: is there agreement between dosimetry methods?

 

Bland-Altman

Correlation and concordance coefficients

\( \Delta {D}_{Mean}^{VOI}(Gy) \)

Bias [95%CI] (Gy)

ρ

ρc

Tumor volume

 PM-MTPM

− 0.7

− 32.5; 31.0

0.94

0.92

 PM-DPK

5.6

− 17.2; 28.5

0.95

0.94

 PM-LDM

6.7

− 15.8; 29.2

0.95

0.93

 MTPM-DPK

6.3

− 12.7; 25.4

1.00

0.96

 MTPM-LDM

7.4

− 14.1; 28.9

0.99

0.94

 LDM-DPK

1.1

− 2.0; 4.1

1.00

1.00

Individual tumors

 PM-MTPM

4.7

− 60.1; 69.5

0.55

0.55

 PM-DPK

7.1

− 52.2; 66.4

0.53

0.50

 PM-LDM

7.9

− 50.1; 65.8

0.53

0.50

 MTPM-DPK

2.4

− 10.6; 15.4

0.99

0.98

 MTPM-LDM

3.2

− 11.8; 18.1

0.99

0.97

 LDM-DPK

0.8

− 2.0; 3.5

1.00

1.00