Fig. 2 | EJNMMI Physics

Fig. 2

From: Performance evaluation of fifth-generation ultra-high-resolution SPECT system with two stationary detectors and multi-pinhole imaging

Fig. 2

a Axial view of tomographic CT images of the three mini Derenzo hot-rod phantoms with different ranges of rod diameters, here shown in mm. All three phantoms were employed for spatial resolution assessment. Phantom 850.500 and 850.100 were also used to create templates for the contrast-to-noise ratio (CNR) analysis. The templates consisting of hot and cold regions of interest (ROI) were placed on the SPECT images of the respective phantoms. Each ROI had a length of 6.0 mm and a diameter of 90% of the corresponding rod diameter. b SPECT image plus CNR template derived from a 45-min scan of the phantom 850.500 acquired with the GP-M collimator and an activity concentration of 30 MBq/ml. The slice thickness was set to 6.0 mm