Table 3 Mean absorbed dose difference (166Ho-DI and 166Ho-only), standard deviation, and lower and upper limits of agreement for VOIs of interest (VOIHEALTHY, VOITUMOR, and VOILUNGS). Top row depicts values for the dataset considered entirely, while middle and bottom rows report values for pre- and post-treatment data, respectively

From: Quantitative 166Ho-microspheres SPECT derived from a dual-isotope acquisition with 99mTc-colloid is clinically feasible

  

VOIHEALTHY

VOITUMOR

VOILUNGS

All data

Average ± SD [Gy]

− 2.00 ± 2.84

− 5.27 ± 8.99

0.80 ± 1.08

Limits of agreement [Gy]

− 7.56 + 3.56

− 22.89

+12.35

− 1.32 + 2.92

Pre-treatment data

Average ± SD [Gy]

− 2.69 ± 3.23

− 4.51 ± 8.69

0.86 ± 1.24

Limits of agreement [Gy]

− 9.02 + 3.64

− 21.54 + 12.52

− 1.57+ 3.29

Post-treatment data

Average ± SD [Gy]

− 1.44 ± 2.27

− 5.96 ± 9.01

0.74 ± 0.92

Limits of agreement [Gy]

− 5.89 + 3.01

− 23.62 + 11.70

− 1.06 + 2.54