Table 1 The measured NEMA NU-2 2018 spatial resolutions averaged between two transaxial planes

From: Performance characteristics of the digital uMI550 PET/CT system according to the NEMA NU2-2018 standard

Radial distance (mm)

FWHM (mm)

FWTM (mm)

Tangential

Radial

Axial

Tangential

Radial

Axial

10

2.91

2.98

2.97

5.35

5.49

5.62

100

3.09

3.29

2.97

5.74

5.83

5.68

200

4.08

4.14

3.13

7.12

7.60

5.91