Table 2 Summary of BSREM β-values that resulted in similar image quality measure, either noise or SNR, as full scan duration OSEM

From: Evaluation of block-sequential regularized expectation maximization reconstruction of 68Ga-DOTATOC, 18F-fluoride, and 11C-acetate whole-body examinations acquired on a digital time-of-flight PET/CT scanner

Image quality parameter

Scan duration (% of OSEM full scan duration)

BSREM β-value

68Ga-DOTATOC

18F-fluoride

11C-acetate

Noise*

100%

400

533

267

Noise

50%

> 533

> 533

400

SNR

100%

200

300

< 133

SNR

50%

400

> 533

133

  1. *Noise was defined as SUVSD divided by SUVmean
  2. BSREM block-sequential regularized expectation maximization, OSEM ordered subsets expectation maximization, SNR signal-to-noise ratio