Skip to main content

Table 1 Quantitative measures of reference sphere in normal liver tissue for 68Ga-DOTATOC, 18F-fluoride, and 11C-acetate

From: Evaluation of block-sequential regularized expectation maximization reconstruction of 68Ga-DOTATOC, 18F-fluoride, and 11C-acetate whole-body examinations acquired on a digital time-of-flight PET/CT scanner

Measure

OSEM

BSREM

β 133

β 267

β 400

β 533

68Ga-DOTATOC

 Volume mL

18.5 (16.8–19.6)

 SUVmax

8.7 (5.1–11.4)

13.5 (9.6–20.1)

9.9 (6.2–13.8)

8.8 (5.1–12.0)

8.3 (4.5–11.5)

 SUVmean

6.2 (3.0–8.0)

6.2 (3.0–8.1)

6.2 (3.0–8.1)

6.2 (3.0–8.1)

6.2 (3.0–8.1)

P*

 

> 0.7

> 0.8

> 0.9

> 0.9

 SUVSD

0.8 (0.4–1.1)

1.5 (0.8–2.0)

1.0 (0.5–1.4)

0.8 (0.4–1.1)

0.7 (0.3–1.0)

 Noise

0.13 (0.07–0.21)

0.25 (0.14–0.44)

0.16 (0.09–0.27)

0.13 (0.08–0.21)

0.12 (0.07–0.18)

18F-Fluoride

 Volume mL

18.5 (16.5–19.6)

 SUVmax

1.0 (0.6–1.3)

3.0 (1.4–6.7)

1.7 (0.8–3.6)

1.1 (0.6–1.5)

0.9 (0.5–1.3)

 SUVmean

0.4 (0.3–0.7)

0.4 (0.3–0.7)

0.4 (0.3–0.7)

0.4 (0.3–0.7)

0.4 (0.3–0.7)

P*

 

> 0.7

> 0.3

> 0.2

> 0.3

 SUVSD

0.1 (0.1–0.2)

0.3 (0.2–0.5)

0.2 (0.1–0.3)

0.1 (0.1–0.2)

0.1 (0.1–0.2)

 Noise

0.30 (0.22–0.40)

0.32 (0.22–0.49)

0.31 (0.16–0.49)

0.33 (0.23–0.49)

0.30 (0.16–0.49)

11C-Acetate

 Volume mL

18.7 (18.0–19.6)

 SUVmax

5.7 (2.5–11.2)

6.3 (2.6–11.3)

5.6 (2.3–10.7)

5.3 (2.2–10.5)

5.2 (2.2–10.3)

 SUVmean

4.3 (1.8–8.2)

4.3 (1.8–8.1)

4.3 (1.8–8.1)

4.3 (1.8–8.1)

4.3 (1.8–8.1)

P*

 

> 0.5

> 0.1

> 0.1

> 0.05

 SUVSD

0.4 (0.2–1.0)

0.5 (0.3–1.1)

0.4 (0.2–0.9)

0.3 (0.2–0.8)

0.3 (0.1–0.8)

 Noise

0.10 (0.07–0.14)

0.13 (0.09–0.16)

0.09 (0.07–0.12)

0.08 (0.05–0.11)

0.07 (0.05–0.11)

  1. †Noise was defined as SUVSD divided by SUVmean
  2. *P: Wilcoxon’s signed rank test, for BSREM to OSEM on SUVmean
  3. BSREM block-sequential regularized expectation maximization, OSEM ordered subsets expectation maximization, SUV standardized uptake value, SD standard deviation