Skip to main content

Table 4 Spatial resolution in a scatter medium for the three probes

From: Comparison of 3 γ-probes for simultaneous iodine-125-seed and technetium-99m breast cancer surgery: NEMA standard characterisation with extended processing

  

FWHM (mm)

FWTM (mm)

Source depth

10 mm

20 mm

30 mm

10 mm

20 mm

30 mm

99mTc

SOE-3211

17

28

40

37

62

91

SOE-3214

20

29

42

38

59

91

GammaSUP-II

15

25

35

30

57

77

125I

SOE-3211

20

30

40

35

56

76

SOE-3214

24

34

45

41

62

84

GammaSUP-II

25

34

48

40

62

86

  1. FWHM full width at half maximum, FWTM full width at tenth maximum