From: Gamma detector dead time correction using Lambert W function
Detector | Energy window | Intrinsic τ (μs) | Point source 1 τ (μs) | Point source 2 τ (μs) | Scatter phantom 1 τ (μs) | Scatter phantom 2 τ (μs) |
---|---|---|---|---|---|---|
Det 1 | OW | 1.30 | 1.05 | 1.07 | 0.92 | 0.94 |
99mTc | 3.16 | 2.69 | 2.61 | 3.6 | 4.29 | |
Det 2 | OW | 1.25 | 1.01 | 0.99 | 0.87 | 0.93 |
99mTc | 3.32 | 2.33 | 2.57 | 3.58 | 4.44 |