Table 2 Dead time values (μs) for the intrinsic point source measurements

From: Gamma detector dead time correction using Lambert W function

Detector

Energy window

τ (μs)

From dual source method

From graphical method

Det 1

OW

1.31 (0.06)

1.30 (0.04)

99mTc

3.19 (0.08)

3.16 (0.04)

Det 2

OW

1.23 (0.04)

1.25 (0.03)

99mTc

3.26 (0.04)

3.32 (0.05)

  1. Dead time values (μs) from the dual source method and the graphical method for the intrinsic point source measurements. Values between brackets are the standard deviation