Table 2 Specifications of the developed human total body PET systems in the Explorer project

From: State of the art in total body PET

System

Penn PET Explorer

uExplorer

Scintillator

LYSO

LYSO

Readout

SiPM (digital)

SiPM

Scintillator size (mm3)

3.86 × 3.86 × 19

2.76 × 2.76 × 18.1

Total number of detector pixels

 

564,480

Bore/detector diameter

70/81

68.6/78.6

Axial length (cm)

70/140

194.8

Resolution (FWHM in mm)

  

Transaxial

4.0

3.0

Axial

4.0

3.0–3.5

Energy res. (%)

10

11.7

Energy window

440-660

430-645

Scatter fraction (%)

NU2-2012

32

35.8

Sensitivity

  

kcps/MBq NU-2

55

191.5 (@0 cm)

Coincidence window (ns)

5 ns

4.5-6.9 ns (ring difference dependent)

TOF resolution (in ps)

250 ps

505 ps

Peak NEC (kcps)

  

NU-2 -2012 (70-cm phantom)

>1200 kcps (for incomplete 70-cm system)

1435 kcps@16.8 kBq/cc

(175-cm phantom)

 

1718 kcps@8.0 kBq/cc