Skip to main content
Fig. 7 | EJNMMI Physics

Fig. 7

From: Monte Carlo simulation of digital photon counting PET

Fig. 7

a Comparison between simulated and experimental images of the relationships between contrast-recovery-coefficients (CRC) and relative noise (RN), according to the number of OSEM iteration and for the 10- and 22-mm diameter hot spheres and the 37-mm cold sphere. b Cross-sectional slices passing through the spheres of the IEC phantom and reconstructed with 2 OSEM iterations and 10 subsets from simulated and experimental data. A parallelepiped profile of 6 mm cross-section and passing through the hot spheres of 10-mm and 22-mm diameter is compared between simulated and experimental images

Back to article page