Skip to main content

Table 1 Systems, acquisition and reconstruction parameters applied in clinical whole-body oncologic 18F-FDG PET procedures

From: Phantom-based image quality assessment of clinical 18F-FDG protocols in digital PET/CT and comparison to conventional PMT-based PET/CT

 

Philips Vereosd

Siemens Visiond

Siemens mCT

GE Discovery-MId,a

GE Discovery 690

System parameters

 Axial ring extent (mm)

164

261

221

250

153

 Energy window (keV)

450–613

435–585

435–650

425–650

425–650

 TOF’s resolution (ps)

316

215

540

370

544.3

 NEMA System sensitivity (kcps/MBq)

5.6

16.4

9.6

22

7.5

Acquistion parameters

 Acq. Time (min)

1.5

2

2.5

2.5

1.5

 Admin. Activity (MBq/kg)

2

2

5

1.5

3.5

 Acq. Time (min) × A admin. (TAP in min × MBq/kg)

3

4

12.5

3.75

5.25

Reconstruction parameters

 Reconstruction methods

OSEM 3D TOF + PSF

OSEM 3D TOF + PSF

OSEM 3D TOF + PSF

OSEM 3D TOF + PSF

OSEM 3D TOF + PSF

 Iterations and subsets (it,ss)

(3,15)/(2,10)

(4,5)

(3,21)

(3,16)

(3,16)

 Filtre Gauss FWHM (mm)

0

0

3

6,4

5

 Matrix size

144 × 144/288 × 288

440 × 440

512 × 512

256 × 256

256 × 256

 Pixel size (mm)

4 × 4/2 × 2

1.65 × 1.65

1.59 × 1.59

2.73 × 2.73

2.73 × 2.73

 Slice thickness (mm)

4/2

2

5

2.79

3.27

  1. aIn addition to OSEM, clinic FDG PET protocol for the GE Discovery MI also make use of the Q.Clear reconstruction algorithm (Q-param = 400)
  2. dDigital PET systems