Skip to main content

Table 5 MAD per system (median and range over all sphere diameters) for data reconstructed using a vendor and center-specific algorithm

From: Towards standardization of absolute SPECT/CT quantification: a multi-center and multi-vendor phantom study

  RCmean RCmax
Discovery NM/CT 670 Pro 0.02 (0.01—0.08) 0.06 (0.02—0.10)
Precendence 6 0.02 (0.00—0.04) 0.03 (0.00—0.06)
Symbia Intevo 6 0.01 (0.01—0.03) 0.01 (0.00—0.05)
Symbia T16 (1) 0.02 (0.00—0.04) 0.02 (0.01—0.05)
Symbia T16 (2) 0.07 (0.00—0.09) 0.09 (0.03—0.19)