Skip to main content

Table 2 Counting rate data

From: Performance characteristics of silicon photomultiplier based 15-cm AFOV TOF PET/CT

Type of measurement Brugge (this work) Stanford [3] Uppsala [3]
Peak NECR (kcps) 102.3 201.1 185.7
Activity at peak NECR (kBq/ml) 23.0 22.1 21.7
Peak true counting ratea (kcps) 463.1 875.9 827.0
Activity at peak true counting ratea (kBq/ml) 36.9 35.4 34.8
Scatter fraction at peak NECR (%) 41.2 40.4 40.8
  1. aNote that the experimental setup did not reach the real peak true counting rate, as explained in the text