Table 4 Contrast, variability, and average lung residual for 8:1 sphere-to-background ratio

From: Performance evaluation of a new high-sensitivity time-of-flight clinical PET/CT system

Sphere Size (mm)

Contrast (%)

Background variability (%)

Non-TOF

Non-TOF-PSF

TOF

TOF-PSF

Non-TOF

Non-TOF-PSF

TOF

TOF-PSF

10

34.3

33.9

41.0

38.2

4.0

4.1

4.6

4.7

13

47.4

48.2

54.2

53.0

3.7

3.8

4.4

4.6

17

60.2

61.5

65.8

64.9

3.4

3.4

4.3

4.4

22

70.3

72.0

74.4

74.4

2.9

3.0

4.3

4.4

28

65.6

62.7

73.0

71.0

2.4

2.6

4.2

4.4

37

70.2

67.6

76.4

74.7

2.0

2.3

4.4

4.6

Lung residual (%)

26.4

27.8

14.9

15.4