Skip to main content

Advertisement

Table 4 Proportion of sites for which reported uncertainty on the measured activity was within the given percentage of the true value. Since H6 did not report an uncertainty, the data from H6 has been excluded from this table

From: Inter-comparison of quantitative imaging of lutetium-177 (177Lu) in European hospitals

  Within 5% Within 10% Within 20% Within 50% Within 75% Within 100%
Inner sphere 3/6 (50%) 5/6 (83%) 6/6 (100%) 6/6 (100%) 6/6 (100% 6/6 (100%)
Outer shell 2/6 (33%) 2/6 (33%) 2/6 (33%) 4/6 (67%) 6/6 (100%) 6/6 (100%)
Total source 4/6 (67%) 5/6 (83%) 6/6 (100%) 6/6 (100%) 6/6 (100%) 6/6 (100%)