Table 3 Proportion of sites for which reported values were within the given percentage of the true value

From: Inter-comparison of quantitative imaging of lutetium-177 (177Lu) in European hospitals

  Within 5% Within 10% Within 20% Within 50% Within 75% Within 100%
Inner sphere 1/7 (14%) 4/7 (57%) 7/7 (100%) 7/7 (100%) 7/7 (100%) 7/7 (100%)
Outer shell 0/6 (0%) 1/6 (17%) 1/6 (17%) 3/6 (50%) 5/6 (83%) 6/6 (100%)
Total source 1/6 (17%) 2/6 (33%) 5/6 (83%) 6/6 (100%) 6/6 (100%) 6/6 (100%)