Table 8 Camera CF obtained using simulated data

From: Determination of gamma camera calibration factors for quantitation of therapeutic radioisotopes

Isotope Configuration A [cps/MBq] Configuration B [cps/MBq] Configuration C [cps/MBq] Configuration D [cps/MBq]
CFPWSCsim CFPPsim \( {\mathrm{CF}}_{R\mathrm{sim}}^B \) \( {\mathrm{CF}}_{R\mathrm{PPsim}}^B \) \( {\mathrm{CF}}_{R\mathrm{sim}}^C \) \( {\mathrm{CF}}_{R\mathrm{PPsim}}^C \) \( {\mathrm{CF}}_{R\mathrm{sim}}^D \) \( {\mathrm{CF}}_{R\mathrm{PPsim}}^D \)
131I 65.74 66.51 69.23 65.55 67.05 67.63 66.54 67.04
177Lu 11.18 11.33 12.44 11.32 11.49 11.59 11.51 11.54
188Re 17.60 18.37 20.47 17.98 18.29 18.77 18.51 18.98