Table 6 Techniques used in CF determination from the experimental and simulated data

From: Determination of gamma camera calibration factors for quantitation of therapeutic radioisotopes

Config.

CF

Definitions

Counts

Time

Activity

Phantom experiments

 A

CFPWSC

Count in PW corrected for scatter using TEW: CPWSC

Scan time: t

Small source activity: A

 B

\( {\mathrm{CF}}_R^B \)

Total counts in the image reconstructed with AC + SC: C R

Number of projections multiplied by the projection duration: n p t p

Total activity in spheres: A

 C

\( {\mathrm{CF}}_R^C \)

Total phantom activity (spheres+bkg): A

 D

\( {\mathrm{CF}}_R^D \)

Total activity in phantom: A

Simulation experiments

 A

CFPWSCsim

Count in PW corrected for scatter using TEW: CPWSCsim

Total number of simulated decays: N tot

CFPPsim

Primary photons simulated in PW: CPPsim

 B

\( {\mathrm{CF}}_{R\mathrm{sim}}^B \)

Total counts in the image that was reconstructed from PW with AC + SC: CRsim

Number of projections multiplied by number of decays simulated in each projection: n p N tot

\( {\mathrm{CF}}_{R\mathrm{PPsim}}^B \)

Total counts in the image reconstructed from primary photons only with AC: CRPPsim

 C

\( {\mathrm{CF}}_{R\mathrm{sim}}^C \)

Total counts in the image that was reconstructed from PW with AC + SC: CRsim

\( {\mathrm{CF}}_{R\mathrm{PPsim}}^C \)

Total counts in the image reconstructed from primary photons only with AC: CRPPsim

 D

\( {\mathrm{CF}}_{R\mathrm{sim}}^D \)

Total counts in the image that was reconstructed from PW with AC + SC: CRsim