Skip to main content

Table 6 Percentage change of SUVmax relative to reference time

From: Initial experience with a SiPM-based PET/CT scanner: influence of acquisition time on image quality

Time (s)

Mean

Standard error

95% confidence interval

30

6.01

2.24

1.48–10.53

60

2.08

1.33

− 0.61–4.76

90

0.95

0.88

− 0.81–2.73

120

0.14

0.63

− 1.12–1.14