Skip to main content

Table 3 Inter-center coefficient of variation (COV) of recovery coefficients (RCs) for each sphere size and activity concentration

From: Multi institutional quantitative phantom study of yttrium-90 PET in PET/MRI: the MR-QUEST study

 

COV (%) (range)

Sphere diameter (mm)

2.16 MBq/ml

1.00 MBq/ml

0.60 MBq/ml

0.36 MBq/ml

0.16 MBq/ml

10

36.2 (14.0–48.6)

86.6 (6.18–99.8)

75.1 (7.79–51.7)

49.5 (9.83–45.1)

64.1 (2.63–26.9)

13

32.2 (23.5–59.8)

55.1 (18.0–68.3)

92.3 (12.5–112)

44.9 (13.6–49.6)

77.5 (3.47–65.5)

17

29.7 (18.9–69.1)

30.4 (30.3–78.0)

36.4 (27.5–75.8)

47.7 (20.5–75.0)

78.8 (14.7–118)

22

22.7 (41.2–84.1)

22.8 (29.6–63.4)

25.0 (32.6–70.0)

23.7 (30.5–56.1)

37.9 (10.4–49.4)

28

24.1 (42.9–88.1)

21.5 (41.7–80.3)

17.8 (38.0–65.3)

26.8 (26.8–69.6)

15.8 (26.8–46.9)

37

18.0 (56.0–95.0)

22.2 (52.0–103)

18.3 (54.9–92.2)

25.3 (46.4–91.4)

21.1 (45.2–81.3)

  1. Values in parentheses represent the range of RCs (min–max)