Skip to main content

Table 1 Clinical scans grading results

From: Feasibility study of a novel general purpose CZT-based digital SPECT camera: initial clinical results

  

Sharpness

Contrast

Uniformity

Overall image quality

Diagnostic confidence

Analog SPECT

Mean

3.26

3.34

3.37

3.66

3.95

Standard deviation

0.50

0.47

1.17

0.55

0.24

Digital SPECT

Mean

3.79

3.92

3.18

3.74

3.79

Standard deviation

0.61

0.29

1.26

0.58

0.50

P value (single-tailed paired T test)

0.0001

< 1e-5

0.1459

0.26

0.0416