Table 3 Summary of depth of interaction determination in three different regions of the detector; the entrance and exit region represent portions of the detector close to the X boundaries while the central part is subtended by the two central SiPMs

From: Depth of interaction determination in monolithic scintillator with double side SiPM readout

Y beam position Investigated region Bias (mm) Z real in [0 mm, 10 mm] Bias (mm) Z real in [2 mm, 8 mm] σ (mm) Z real in [0 mm, 10 mm]
0 mm Entrance
X > 3.6 mm
0.20 0.21 1.33
Central region 0.21 0.18 1.54
Exit
X < −3.6 mm
0.25 0.21 1.60
1.8 mm Entrance
X > 3.6 mm
0.19 0.22 1.39
Central Region 0.18 0.11 1.46
Exit
X < −3.6 mm
0.41 0.46 1.60