Table 3 Details of planar and tomographic acquisitions performed to determine the CNF for 177Lu

From: Accuracy of 177Lu activity quantification in SPECT imaging: a phantom study

 

Mode (phantom)

Fig. 5 data

Phantom

Source info

Acquisition parameters

A

Planara (point source)

Planar methods 1 and 2

Activity 11.7 MBq

Volume 0.5 ml

Scan duration = 10 min

Collimator MELP

PW [187.2–228.8] keV

LSW [153.0–187.0] keV

USW [229.5–280.2] keV

B

Tomographicb (uniformly hot cylinder)

SPECT cylinder

Total activity 659.6 MBq

Volume 6500 ml

#Projections = 96

Time/projection = 10s

C

Tomographic (6 bottles in air)

SPECT air

Total activity 233.4 MBq

Vol bottles [17–199.5] ml

#Projections = 90

Time/projection = 20s

D

Tomographic (6 spheres in warm water)

SPECT HW1

Total activity 489.1 MBq

Vol bottles [17–199.5] ml

#Projections = 90

Time/projection = 30s

E

Tomographic (6 spheres in warm water)

SPECT HW2

Total activity 681.3 MBq

Vol bottles [17–199.5] ml

#Projections = 90

Time/projection = 30s

  1. aThe camera sensitivity does not depend on the source-collimator distance although increase in recorded counts may occur due to septal penetration if the sources are placed very close to the collimator. To minimize this effect, in this study the point sources were placed at 30 cm from the collimator surface
  2. bUniform cylinder was scanned at the L’Hôtel-Dieu de Québec site of the CHU de Québec–Université Laval center (Quebec City, Canada) and dual energy window was performed for scatter correction on this phantom