Skip to main content

Table 2 Contrast results (in percent) for the NEMA phantom with various sphere sizes, with different reconstruction settings and the 1:8 ratio for the Biograph mMR and Biograph mCT

From: Quantitative comparison of PET performance—Siemens Biograph mCT and mMR

 

mMR

mCT

OSEM

PSF

OSEM

PSF

TOF

PSF + TOF

Inner diameter (mm)

      

 10

36.1

39.2

35.1

41.4

36.2

48.2

 13

62.1

70.2

56.7

68.8

57.1

70.3

 17

73.9

79.2

64.3

72.5

67.7

74.3

 22

80.5

83.0

74.2

78.3

76.7

80.7

Mean (hot spheres)

63.1

67.9

57.6

65.2

59.4

68.4

 28

60.9

60.5

71.7

71.3

79.1

80.1

 37

67.3

67.7

77.6

77.5

84.9

85.5

Mean (cold spheres)

64.1

64.1

74.7

74.4

82.0

82.8