Table 1 Spatial resolution measured for the PET component of the mCT Flow system (NEMA NU2-2012) and comparison with the published values (NEMA NU2-2007) of its predecessor, the mCT system [6]

From: Performance evaluation of the Biograph mCT Flow PET/CT system according to the NEMA NU2-2012 standard

Spatial resolution

Distance (cm)

Measured (mCT Flow)

Published (mCT) [6]

FWHM (mm)

FWTM (mm)

FWHM (mm)

FWTM (mm)

Transverse

1

4.33

8.60

4.4 ± 0.1

8.6 ± 0.1

Axial

1

4.25

8.55

4.4 ± 0.1

8.7 ± 0.2

Transverse radial

10

5.16

9.30

5.2 ± 0.1

9.4 ± 0.1

Transverse tangential

10

4.72

9.68

4.7 ± 0.1

9.2 ± 0.1

Axial

10

5.85

11.06

5.9 ± 0.1

10.9 ± 0.3

Transverse radial

20

5.55

9.84

Transverse tangential

20

6.48

12.68

Axial

20

7.80

13.7