Table 3 Contrast recovery and background variability values in % for each active sphere for different reconstruction parameters of the 3D OP-OSEM reconstruction algorithm when using CT-based AC

From: NEMA image quality phantom measurements and attenuation correction in integrated PET/MR hybrid imaging

 

aContrast recovery [%] │ aBackground variability [%]

 

Sphere 10 mm

Sphere 13 mm

Sphere 17 mm

Sphere 22 mm

Iteration

172 x 172

344 x 344

172 x 172

344 x 344

172 x 172

344 x 344

172 x 172

344 x 344

1

13.8 │ 2.5

15.8 │ 2.8

28.1 │ 2.3

29.6 │ 2.5

48.5 │ 2.1

51.3 │ 2.3

53.7 │ 2.1

56.0 │ 2.2

2

24.3 │ 3.2

28.1 │ 3.5

43.8 │ 2.6

45.3 │ 2.8

66.8 │ 2.3

69.7 │ 2.3

70.0 │ 1.9

72.0 │ 2.2

3

30.5 │ 3.8

35.0 │ 4.3

50.5 │ 3.0

51.1 │ 3.2

72.9 │ 2.5

75.2 │ 2.6

74.5 │ 2.3

76.3 │ 2.4

4

34.3 │ 4.4

38.7 │ 4.9

53.4 │ 3.3

53.3 │ 3.5

75.4 │ 2.7

77.5 │ 2.8

76.5 │ 2.5

78.2 │ 2.6

5

36.8 │ 4.8

40.8 │ 5.3

54.8 │ 3.5

54.3 │ 3.8

76.9 │ 2.9

78.7 │ 3.0

77.6 │ 2.6

79.2 │ 2.7

  1. a4 mm Gaussian filtering and 21 subsets were kept constant for all reconstructions. The values represent the mean of four independent measurements